When I use my Rapid RH® Easy Reader, all I see are 2 lines. What should I do?

If the reader is displaying two lines (–) then immediately back it out of a Smart Sensor. Check to see if this problem occurs with multiple sensors (not just related to an individual sensor). If the problem occurs with another Smart Sensor, then most likely the problem is that the connector at the bottom of the Easy Reader is not making proper contact with the sensor(s). Inspect the connector to see if any of the leaf springs is damaged, bent, twisted, etc. If that does not result in a solution, make sure you have tried a new set of batteries. If the problem still exists, the proper recourse is to send it in for repair and/or replacement.

Please visit www.wagnermeters.com for more information.

The following two tabs change content below.
Wagner Meters is a family-owned American business that aims to provide solutions in moisture measurement technology that will enhance the quality and value of each customer's project. With an almost 60-year legacy of innovation, Wagner continues to be a resource for both individual craftsmen and high-performance commercial endeavors.